Ultimate Knowledge | Calibration method of temperature and humidity calibration chamber

2024-2-5 This article is read 5057 Times

DTLH copywriting.jpg

1. Sensor location

Before calibration, the effective working area shall  be  determined  according to the  operation manual of the temperature and humidity chamber or the user   requirements.Multi-channel temperature and humidity measuring device is used to calibrate the temperature and humidity in the chamber. The temperature sensor and humidity sensor in the device adopt horizontal layered layout. The location from the side of the effective working area is 8 cm. The lower location is 5 cm above the bottom level of the effective working area, and the upper location is 20 cm below the top level of the

effective working area, as shown in Figure 1.

Temperature and humidity chamber with the effective working area height,h,less than 50 cm,

locate 5 points [see Figure la)

Temperature and humidity chamber with the effective working area height ,h, between 50 c m

and 100 c m, locate 9 points [see Figure lb)

Temperature and humidity chamber with the height of the effective working area,h,exceeding

100 cm, locate 13 points [see Figure lc)].

a)   h <50 cm distribution plot

b) 50 cm ≤ h ≤100 cm  distribution plot


c) h> 100 cm distribution plot

Figure 1 Schematic diagram of the sensor spot placement

For  the  temperature and humidity  chamber  with  input cold mirror dew point probe in  the working room, the distribution position of the sensor should be adjusted appropriately, so that the distance between the sensor in the upstream of the dew point probe and the air flow in the chamber is not less than 5 cm, and the distance between the sensor downstream of the dew point probe and the

dew point probe is not less than 20 cm.

For the temperature and humidity chamber with airflow blind area in the effective working area, a distribution point can be added in the blind area.


2. Temperature calibration

2. 1 Temperature calibration point

Temperature calibration points are generally 15 °C, 20 °C, and 30 °C, and other calibration points

can also be selected according to user requirements.During the calibration, the temperature shall be conducted in the order of low to high, and the temperature deviation at each calibration point shall be within  ±0.5 °C (whichever is the center point temperature).If the humidity function is turned on, the humidity  set  value  is  generally  about  60%RH.The  calibration  items  are  generally  conducted  under no-load conditions.

2. 2 Temperature calibration steps

After   the   temperature   and    humidity   chamber   reaches   the   specified    stability   time,   the temperature  of  each  point  is  recorded.  The  recording  interval  of  the  center  point  is  20s,  and  the recording interval of the other points is 2 min.Center points recorded 91 sets of data and other points recorded 16 sets of data.20s interval data use T ’i0  (i =  1,2, ??? ,91) indicates that 2 min interval data were used with Tnj (n=l, 2, ??? , 16; j =0,1, ??? , m-1), where m is the number of points, i and n represents the serial  number  of time  (i  time  interval  is  20s,n  time  interval  is  2  min),  j  representing  the  serial number of point position.If the temperature and humidity in the chamber can be determined to be

balanced before the specified stability time, it can also be recorded in advance.

The temperature stability time shall be subject to the instructions, if not given in the instructions, the following principles: if the temperature change does not exceed 10 °C, the stability time from the temperature and humidity setting to the start recording data is 90 min; if the temperature change does not exceed 5 °C, the stability time is 60 min.The stabilization times of each calibration point are shown

in Table 1.

Table 1 Stability Time



Ultimate Knowledge | Calibration method of temperature and humidity calibration chamber
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:New Product | 180C to 670C temperature calibration intelligent precision thermostatic salt bathnext:Invitation: Join Us at World Metrology Day 2026 Conference in Singapore

Copyright? Tai'an Dearto Automation Instrument Co., Ltd.

Chinese

主站蜘蛛池模板: а中文在线天堂 | 香蕉在线影院 | 波多在线视频 | 在线а√天堂中文官网 | 久久精品国产免费看久久精品 | 人人插人人舔 | 天天干,天天草 | 国内精品久久久精品电影院 | 区一区二在线 | 高清一区二区三区av | 日本黄色免费看 | 91久久黄色 | 国产精品视频地址 | 一级黄毛片 | 日本爱爱免费 | 精品国产乱码一区二 | 狠狠操91 | 国产精品美女毛片真酒店 | 夜夜爽夜夜操 | 久久夜色精品亚洲噜噜国4 午夜视频在线观看欧美 | 涩涩网站免费 | 婷婷丁香视频 | 日本xxxx裸体xxxx17 | 久草在线最新免费 | ww视频在线观看 | 亚洲免费在线看 | 区一区二在线 | 久久免费美女视频 | 中文字幕精品在线 | 久久成人黄色 | 国产 亚洲 欧美 在线 | 亚洲精品午夜一区人人爽 | 国内精品久久久久久久97牛牛 | 亚洲成年人在线播放 | 日本公妇在线观看 | 亚洲成人黄色av | 91激情视频在线播放 | 国产精品久久久久免费 | 91高清不卡| 韩国av在线播放 | 色婷婷视频 | 欧洲亚洲国产视频 | 亚洲 欧美日韩 国产 中文 | 国产美女精品视频免费观看 | 91精品视频免费在线观看 | 国产成人精品亚洲日本在线观看 | 成年人天堂com | a视频免费 | 香蕉视频在线看 | 在线一区电影 | 国产最新在线 | 国内小视频在线观看 | 久久五月天婷婷 | 成人午夜精品福利免费 | 国产精品网站一区二区三区 | 国产一区在线播放 | 丁香五香天综合情 | 色搞搞| 免费观看久久久 | 国产白浆视频 | 久草在线一免费新视频 | 97福利在线 | 久久亚洲影视 | 国产成人精品在线观看 | 射射射av| 免费麻豆视频 | 中文字幕国产视频 | 精品国产精品久久一区免费式 | 亚洲精品欧美精品 | 亚洲精品高清一区二区三区四区 | 黄色三级久久 | 免费午夜网站 | 丁香婷婷网 | 欧亚日韩精品一区二区在线 | 日日草夜夜操 | 夜添久久精品亚洲国产精品 | 中文乱码视频在线观看 | 韩日精品在线 | 亚洲精品综合一二三区在线观看 | 久久久久久综合网天天 | 五月婷婷丁香 | 国产 色 | 欧美男男tv网站 | 亚洲国内精品 | 五月综合婷 | 国产精品18久久久久久久久 | 天堂网在线视频 | 日韩精品久久久免费观看夜色 | 国产123av | 久久大香线蕉app | 香蕉影视app| 精品一二三区视频 | 在线一区观看 | 久久精品日产第一区二区三区乱码 | 日本中文字幕网站 | 人人狠 | 中文字幕日韩av | 国产日韩视频在线观看 | 亚洲国产激情 |