Ultimate Knowledge | Calibration method of temperature and humidity calibration chamber

2024-2-5 This article is read 5074 Times

DTLH copywriting.jpg

1. Sensor location

Before calibration, the effective working area shall  be  determined  according to the  operation manual of the temperature and humidity chamber or the user   requirements.Multi-channel temperature and humidity measuring device is used to calibrate the temperature and humidity in the chamber. The temperature sensor and humidity sensor in the device adopt horizontal layered layout. The location from the side of the effective working area is 8 cm. The lower location is 5 cm above the bottom level of the effective working area, and the upper location is 20 cm below the top level of the

effective working area, as shown in Figure 1.

Temperature and humidity chamber with the effective working area height,h,less than 50 cm,

locate 5 points [see Figure la)

Temperature and humidity chamber with the effective working area height ,h, between 50 c m

and 100 c m, locate 9 points [see Figure lb)

Temperature and humidity chamber with the height of the effective working area,h,exceeding

100 cm, locate 13 points [see Figure lc)].

a)   h <50 cm distribution plot

b) 50 cm ≤ h ≤100 cm  distribution plot


c) h> 100 cm distribution plot

Figure 1 Schematic diagram of the sensor spot placement

For  the  temperature and humidity  chamber  with  input cold mirror dew point probe in  the working room, the distribution position of the sensor should be adjusted appropriately, so that the distance between the sensor in the upstream of the dew point probe and the air flow in the chamber is not less than 5 cm, and the distance between the sensor downstream of the dew point probe and the

dew point probe is not less than 20 cm.

For the temperature and humidity chamber with airflow blind area in the effective working area, a distribution point can be added in the blind area.


2. Temperature calibration

2. 1 Temperature calibration point

Temperature calibration points are generally 15 °C, 20 °C, and 30 °C, and other calibration points

can also be selected according to user requirements.During the calibration, the temperature shall be conducted in the order of low to high, and the temperature deviation at each calibration point shall be within  ±0.5 °C (whichever is the center point temperature).If the humidity function is turned on, the humidity  set  value  is  generally  about  60%RH.The  calibration  items  are  generally  conducted  under no-load conditions.

2. 2 Temperature calibration steps

After   the   temperature   and    humidity   chamber   reaches   the   specified    stability   time,   the temperature  of  each  point  is  recorded.  The  recording  interval  of  the  center  point  is  20s,  and  the recording interval of the other points is 2 min.Center points recorded 91 sets of data and other points recorded 16 sets of data.20s interval data use T ’i0  (i =  1,2, ??? ,91) indicates that 2 min interval data were used with Tnj (n=l, 2, ??? , 16; j =0,1, ??? , m-1), where m is the number of points, i and n represents the serial  number  of time  (i  time  interval  is  20s,n  time  interval  is  2  min),  j  representing  the  serial number of point position.If the temperature and humidity in the chamber can be determined to be

balanced before the specified stability time, it can also be recorded in advance.

The temperature stability time shall be subject to the instructions, if not given in the instructions, the following principles: if the temperature change does not exceed 10 °C, the stability time from the temperature and humidity setting to the start recording data is 90 min; if the temperature change does not exceed 5 °C, the stability time is 60 min.The stabilization times of each calibration point are shown

in Table 1.

Table 1 Stability Time



Ultimate Knowledge | Calibration method of temperature and humidity calibration chamber
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:New Product | 180C to 670C temperature calibration intelligent precision thermostatic salt bathnext:Invitation: Join Us at World Metrology Day 2026 Conference in Singapore

Copyright? Tai'an Dearto Automation Instrument Co., Ltd.

Chinese

主站蜘蛛池模板: 亚洲视频电影在线 | 香蕉视频在线视频 | 亚洲天堂精品 | 国产91粉嫩白浆在线观看 | 免费人做人爱www的视 | 狠狠色丁香婷婷综合久小说久 | 操少妇视频 | 国产黄免费看 | 精品一区二区在线播放 | 久久伊人八月婷婷综合激情 | 婷婷六月网 | 91九色九色| 黄色av电影在线观看 | 狠狠狠干狠狠 | 色资源网免费观看视频 | 国产欧美久久久精品影院 | 日韩有码网站 | 久久艹在线 | 国产亚洲综合性久久久影院 | 亚洲精品三级 | 婷婷国产在线 | 91av视频导航 | 精品久久久免费视频 | 成片视频免费观看 | 亚洲狠狠婷婷综合久久久 | 精品国产一区二区三区不卡 | 免费福利在线 | 天天综合久久综合 | 免费欧美精品 | 91精品老司机久久一区啪 | 日韩电影一区二区在线观看 | 久久人91精品久久久久久不卡 | 波多野结衣电影一区 | 久久黄色影院 | 一区二区三区在线观看免费 | 中文字幕视频播放 | 久久www免费人成看片高清 | 深爱激情五月婷婷 | 福利电影久久 | 成人久久久精品国产乱码一区二区 | 国产黄色精品在线 | 午夜久久美女 | 国产高清成人在线 | 色中文字幕在线观看 | 欧美激情精品久久久久久免费印度 | 三级黄色在线观看 | 中文字幕在线国产精品 | 久久精品欧美一区二区三区麻豆 | 丁香婷婷久久久综合精品国产 | 天天干天天操天天射 | 国产成人中文字幕 | 精品免费久久久久 | 毛片激情永久免费 | 国产免费观看久久 | 亚洲综合色播 | 国产黄色大片 | 又黄又爽的免费高潮视频 | 麻豆视频免费看 | 国产色资源 | 日韩激情片在线观看 | 一区精品在线 | 97精品久久人人爽人人爽 | 综合色影院| 最近最新中文字幕 | 久久国产女人 | 国产精品福利午夜在线观看 | 一区二区三区手机在线观看 | 久热这里有精品 | 成人午夜电影在线观看 | 国产精品成久久久久 | 少妇bbbb | 久久免费国产 | 国产精品久久在线观看 | 丁香五月网久久综合 | 中文字幕 国产精品 | 在线视频1卡二卡三卡 | 久久久久女人精品毛片九一 | 婷婷六月综合亚洲 | 国产美女免费观看 | 久久精品日本啪啪涩涩 | 日韩欧美精品免费 | 亚洲另类在线视频 | 色婷婷av在线 | 欧美 亚洲 另类 激情 另类 | 91av在线免费 | 五月婷亚洲 | 青青色影院 | 日韩精品一区二区三区免费观看视频 | 在线免费91 | 91最新网址在线观看 | 日本视频网 | 亚洲高清久久久 | 日韩理论在线视频 | 人人看看人人 | 992tv在线观看网站 | 久久久久亚洲国产精品 | 天天综合色网 | 午夜在线观看影院 | 国产不卡av在线 |