Precision Refrigeration Calibration Baths PC Operation Method

2022-12-8 This article is read 5693 Times

Precision Refrigeration Calibration Baths PC Operation Method

Precision Calibration Bath System overview

The touch screen precision refrigeration calibration bath is a new product controlled by an intelligent touch screen operating system. The system can automatically start and stop temperature control, segmental temperature correction and linear rate temperature rise and fall. Users can directly operate through the touch screen on the calibration bath, which is more convenient and intelligent.

The main functions of the touch screen calibration slot are:

1. Automatic temperature control: The system automatically performs heating, stirring and cooling control according to the calibration operation status, without manual operation.

2. Multi-point calibration: The system supports segmental temperature correction, and automatically completes temperature control according to the correction value during temperature control, avoiding repeated correction when the temperature rises.

3. Linear temperature control: The system supports linear temperature rise and fall according to the specified rate to meet the temperature control requirements.

4. Multi-point setting: The system supports setting at different temperature points and automatically saves the setting parameters. When controlling the temperature, it automatically controls the temperature according to the setting parameters to ensure accurate temperature control at any point.

5. Overheating protection: The system supports overheating protection. When the temperature is overheated, it will automatically stop heating to protect the equipment.

6. Remote communication: The system supports remote communication control through serial port, WLAN, WIFI.

7. Mobile phone control: The system supports device control through the mobile phone App.

Precision Calibration Bath Interface display

After the system is started, it is in standby mode. Click the button  to start the system and start automatic temperature control.

The upper status bar displays the task animation according to the actual operation situation,  indicating that heating is in progress,;indicating that stirring is in progressindicating that cooling 1 is on and that cooling 2 is on.

The precision calibration bath system is divided into two cooling modes. When the automatic cooling mode is enabled, the system automatically controls the logic of heating, stirring, and cooling. At this time, when the corresponding task is enabled, the task animation is displayed, and when the corresponding task is disabled, the task animation is hidden.

Precision Refrigeration Calibration Baths PC Operation Method

When the automatic cooling mode is deactivated, the system enters the manual cooling mode, and the corresponding tasks need to be controlled manually. When a task is disabled, a gray dormant task icon is displayed. When the task is started manually, the task animation is displayed.

Precision Refrigeration Calibration Baths PC Operation Method

When the set fluctuation statistics time is reached, the temperature fluctuation will be displayed in real time.

When the temperature is stable, the temperature stabilization time will be displayed in real time.

Precision Refrigeration Calibration Baths PC Operation Method

Click the set value temperature7.jpg to open the temperature setting interface, enter the set value, and complete the modification of the set temperature.

Precision Calibration Bath Curve Display

Click on the main interface to enter the curve display interface8.jpg. The curve display interface will automatically update the view according to the collected data, and display the curves of all data in real time.

Zooming and panning operations can be performed when the curve is displayed. After the operation, it enters the manual control mode and is displayed in the upper right corner16705509446362119.jpg. At this time, the view is no longer automatically updated. Click16705509446362119.jpg.  , the system exits the manual mode, the curve interface returns to the automatic control mode, and automatically updates the view according to the collected data.

Pan operation: Press and hold the curve with a single finger to drag up, down, left, and right, and the curve moves with the finger operation.

Zoom operation: hold down the screen with two fingers and move in opposite directions, the curve will zoom in with the gesture. Moving in the same direction, the curve shrinks with the gesture.

Click the button to exit the curve interface and return to the main interface.

Precision Refrigeration Calibration Baths PC Operation Method

Precision Calibration Bath Configuration Interface

Click on the main interfaceto pop up the function menu, clickto enter the configuration interface.

Temperature unit: After modifying the temperature unit, the main interface will display the temperature value according to the newly set unit.

Precision Refrigeration Calibration Baths PC Operation Method

Resolution: After modifying the resolution, the main interface will display the temperature according to the newly set resolution.

Precision Refrigeration Calibration Baths PC Operation Method

Volatility: The volatility setting is used to determine the statistical cycle of the volatility and whether the volatility is displayed on the main interface.

Statistical period: After modifying the statistical period, the calibration bath system calculates the difference between the highest temperature and the lowest temperature in the statistical period according to the new statistical period as the fluctuation degree.

Enable prompt: When enabled, after the system running time exceeds the statistical cycle time, the fluctuation degree will be displayed in real time on the main interface automatically. When closed, the main interface no longer displays volatility.

Precision Refrigeration Calibration Baths PC Operation Method

Stability prompt: Stability prompt configuration is used to determine the statistical parameters of temperature stability and whether the temperature stability is displayed on the main interface.

Statistical period: After modifying the statistical period, the system calculates the temperature fluctuation for statistics according to the new statistical period. The statistical period for temperature stability may be different from the statistical period for volatility.

Stable fluctuation, stable deviation: When the calculated stable statistical fluctuation is less than the set stable fluctuation and the calculated temperature deviation is less than the stable deviation, the system enters a stable state and begins to count the stable time.

Stability monitoring: After the system enters a stable state, it will always remain in a stable state until the calculated fluctuation of the stable statistics is greater than the set stability monitoring, precision calibration bath system exits the stable state, the stabilization time is reset, and the stable statistics need to be restarted.

Enable prompt: When enabled, the real-time temperature stabilization time will be automatically displayed on the main interface after the system determines that it has entered a stable state. When closed, the main interface no longer displays the temperature stabilization time.

Precision Refrigeration Calibration Baths PC Operation Method

Serial port communication: The serial port communication configuration determines whether to open the serial port service. After opening, the client can control the temperature calibration through the serial port communication. The communication speed of the serial communication is 9600, the data bit is 8, the parity bit is N, and the stop bit is 1.

Precision Refrigeration Calibration Baths PC Operation Method

Network communication: The network communication configuration determines whether to open the network service. After opening, the client can control the oil tank through network communication, or control the oil tank through the mobile phone client.

Precision Refrigeration Calibration Baths PC Operation Method

Precision Calibration Bath Advanced configuration

The advanced configuration provides various configuration options such as temperature correction, cooling mode, cooling setting, acquisition setting, PID parameters, temperature control rate, overheating protection, etc., and provides finer control of the cooling calibration bath, and the user does not need to make adjustments during use. If operation is required, please contact the manufacturer.

Precision Refrigeration Calibration Baths PC Operation Method

Precision Refrigeration Calibration Baths PC Operation Method
  • Online message

  • Your name
    Your title must be filled
  • Cellphone number
    Mobile phone number is required
  • whatsapp/wechat
    Optional
  • Type of consultation
    Required, please select the type of consultation
  • Consultation content
    Required, what is the content you want to consult?
  • Verification code: Please fill in the calculation result.
Previous:How do you Calibrate a Bath Temperature?next:Invitation: Join Us at World Metrology Day 2026 Conference in Singapore

Copyright? Tai'an Dearto Automation Instrument Co., Ltd.

Chinese

主站蜘蛛池模板: 久久久久电影 | 久久理论电影 | 丁香av在线| 开心激情网五月天 | 欧美 亚洲 另类 激情 另类 | 亚洲电影第一页av | 国产精品国产三级国产专区53 | 欧洲精品视频一区 | 欧美激情精品久久久久久变态 | 中文在线www | 免费观看第二部31集 | 正在播放一区 | 日本一区二区三区免费看 | 91九色视频在线 | 在线观看蜜桃视频 | 99性视频 | 欧美亚洲久久 | 成人免费看视频 | 一区二区三区视频网站 | 国产高清av免费在线观看 | 天天爱天天射天天干天天 | 中文字幕888| 亚洲精品视频网站在线观看 | 欧美国产不卡 | 91av精品| 成人资源网 | 精品一区av| 444av| 一区二区三区精品久久久 | 国产精品久久久久av福利动漫 | 日韩欧美极品 | 欧美一二三四在线 | 国产精品成人自产拍在线观看 | 久久不见久久见免费影院 | 在线99 | 天堂成人在线 | 在线观看视频一区二区 | 最近字幕在线观看第一季 | 97超碰色 | 五月激情视频 | a在线免费观看视频 | 在线国产高清 | 丁香婷婷色综合亚洲电影 | 婷婷丁香av | 久久免费视频1 | 欧美另类交人妖 | 波多野结衣在线观看一区二区三区 | 日韩免费精品 | 亚洲欧美日韩国产 | 亚洲综合成人婷婷小说 | 五月婷婷国产 | 免费在线黄色av | 一区二区三区免费在线 | 尤物九九久久国产精品的分类 | 免费三级a | 欧美性受极品xxxx喷水 | 91精品啪啪| 久久99精品久久只有精品 | av播放在线 | 亚洲视频六区 | 色噜噜日韩精品一区二区三区视频 | www.亚洲精品视频 | 欧美日韩不卡一区 | 国产在线观看,日本 | 国产中文在线视频 | 三级在线国产 | 色大片免费看 | 人人爱天天操 | 三级在线视频观看 | 中文字幕在线观看你懂的 | 欧美成人xxxxx| av色一区 | 国产精品一区二区在线播放 | 国产精品免费一区二区三区在线观看 | 99热这里只有精品久久 | 99国产一区 | 92国产精品久久久久首页 | 99久国产 | 久久久在线观看 | 久久精品99国产精品亚洲最刺激 | 婷婷激情五月综合 | 国产精品mm| 在线网站黄 | 日本在线观看一区二区三区 | 国产免码va在线观看免费 | 五月天综合在线 | 五月天天av| 欧美亚洲一区二区在线 | 欧美在线日韩在线 | 色噜噜日韩精品一区二区三区视频 | 欧美大片mv免费 | 日韩经典一区二区三区 | 99热日本| 91在线91拍拍在线91 | 久久久久激情视频 | 美女久久久久久 | 中文字幕影片免费在线观看 | 91视频在线自拍 | 九九在线视频免费观看 |